index  
coursenewsroomcompanycontactlogin
 
    Product Details
  Indications for Use/Contraindications
  Surgical Technique
   
      Bone Preparation
      Screw Insertion

Patient Positioning/Positioning Guide Pin/
Screw Length Measurement
Previous | Next Next


 

Patient Positioning

The LARGE QWIX® screws and instrumentation are color coded :

    Yellow for 5.5 mm diameter screws
    Purple for 7.5 mm diameter screws


    Positioning Guide Pin

  • A 1.6 or 2.5mm K-wire should be introduced to achieve temporary fixation and maintain alignment.
    A drill guide and soft tissue protector is offered to prevent entrapment of the soft tissues during insertion (figure 1).


  • (Note: The use of x-rays is recommended to check the positioning and depth of the K-wire.)

    Screw Length Measurement

  • The measuring device is graduated in accordance with the various K-wires included within the set. The measuring device should be placed along the
    K-wire such that the tip is in contact with the surgical site. The appropriate screw length is determined directly from the measuring device on the side marked. (K-wire 200mm/ K-wire 250mm) (figure 2)

    (Note: The measurement taken from the measuring device is directly related to the positioning of the guide pin and therefore, is not considered accurate unless the guide pin is positioned correctly. The surgeon may decide to add or deduct a few millimeters depending on the position of the guide pin and the desire to obtain mono or bi-cortical fixation.)


 


 
Previous | Next Next

© Copyright 2010 Integra LifeSciences Corporation. All Rights Reserved.
Terms of Use • Privacy Policy