    Sizer Insertion

  • Place the cannulated, 6mm sizer over the guide pin and insert it through the sinus tarsi into the sinus canalis from lateral to medial (Figure 5)
  • Assess the range of motion of the subtalar joint
  • Continue to insert the remaining sizers (8, 9, 10, 12mm) until proper correction is achieved
  • The appropriate size will allow approximately 2-4 degrees of subtalar joint eversion

    Trial Implant Insertion

  • After the appropriate size is determined, remove the sizer
  • Insert the corresponding Trial Implant using the cannulated 3.5mm Hex Insertion Tool (Figure 6)
  • Assess the range of motion of the subtalar joint
  • Use intra-operative imaging to evaluate the degree of correction and placement of the Trial Implant

 