Once the size of the endorthesis is defined,
the implant is fixed on the specific Kalix®
holder by turning the holder clockwise
The peaks of the unscrewer will be introduced
in the imprints of the metallic insert for the
unscrewing maneuver
The endorthesis is set in the Sinus Tarsi,
at the level of the lateral border of the talus
(10)
The holder is retrieved from the implant by
unscrewing (counter-clockwise) of the holder while
the unscrewer remains in steady position (with
the two peaks in the metallic cone)
The double screwdriver is used for the final
implantation of the endorthesis (holding it in
the sinus tarsi along with a press-fit positioning)
The external part of the screwdriver has two
peaks which fit into the imprints in the metallic
conical insert, whereas the central part of the
screwdriver will fit into the central axis of
the endorthesis
The final setting of the implant is achieved
by clockwise manipulation of the internal part
of the screwdriver
The screwing maneuver will be completed by the
locking of the system.