  Mid & HindFoot Solutions
    Product Details
    Indications for Use/Contraindications
   
      Pre-Operative Planning
      Calcaneal Reconstruction
      Calcanea® Plate Positioning
   
      Calcanea® Plate Fixation using Variable Angle Screws
      Post-Operative Care
 

Calcanea® - Surgical Technique
Previous | Next Next
  • The drilling sleeves (129135ND) are first screwed in the hole located
    on the anterior extremity and in the 2 holes on the posterior extremity of the plate
  • They allow axial drilling and perfect insertion of the locking screws
  • The holes for fixation screws are drilled using the 2.2 mm drill (119006ND)
    through the drilling sleeves

  • The plate is fixed using specific 3.5 mm screws
  • The appropriate length of the screws to be inserted is evaluated
    using the depth gauge (129134ND)
  • Color coding of each size of screw allows for quick choice of the
    adequate length to be used

  • Holes in the plate allow fixation with up to 4 screws in the posterior
    tuberosity (2 locking screws and 2 variable angle screws), and 3 screws
    in the anterior process (1 locking screw and 2 variable angle screws)
  • If there is a fracture of the anterior process of the calcaneus, it will be
    temporarily stabilized with Kirschner wires
  • The screws are inserted with the specific screwdriver (129132ND)
    in a normal way, and introduced in the bone until the base of their
    head is blocked against the plate

 
Previous | Next Next