    Compressing the Uni-CP™ Plate

  • After all screws are locked in place, compress the Uni-CP™ using the desired spreading instrument. Upon opening the diamond shaped bridge, the compressive forces will pull the ends of the plate toward one another (figure 6).

    Note: Large plates require more force and can be difficult to open using the small instrument. Small plates can easily be over-compressed when using the large spreading instrument.
 
 
 