Indications for Use/Contraindications
   
      Articular Surfaces Reconstruction
      Trial Plate Positioning, Plate Contouring, Implant Positioning
      Drill Screw Holes, Screw Insertion
   
      Tibial Fixation
      Talar Fixation
 

Joint Space Reduction
Previous | Next Next
  • After insertion of the calcaneal screws, compression is applied and maintained to
    reduce the arthrodesis (figure 7a)

  • A drill guide (219 645ND) is fixed to the most proximal hole (figure 7b)
  • A 2.5mm K-wire (115 425ND) is inserted into the tibial diaphysis, 20mm from the plate
  • The compression clamp (219 900ND) is fixed, in a open position, to the drilling guide and the K-wire
  • Temporary K-wires are removed

  • Compression is applied with the compression
    clamp (219 900ND) as shown in figure 7c


 
 