index  
coursenewsroomcompanycontactlogin
 
 
  Mid & HindFoot Solutions
  Product Details
  Indications for Use/Contraindications
  Surgical Technique (4.0 mm)
 
    Positioning the Piin
    Determining Length of Screw
    Use of Drill
    Use of the Tap
    Screw Insertion
 
    Optional Translation of Screw Body

Surgical Technique (6.5 mm)
Previous | Next Next

Optional Dynamic Compression

External Screwdriver 119566ND 6.5 mm

  • The static compression effect can be increased by continuous screwing movement of the entire screw. However, in order to maintain effective compression and adequate stability the head of the screw should remain in the cortical bone.
  • Additional (optional) dynamic compression is achieved by advancing the outer thread of the screw independently of its main body
  • For this, the external screwdriver (119566ND 6.5 mm) is used, which screws the external threading of the screw head. The body (internal part) of the screw will not be moved.

Previous | Next Next

© Copyright 2005 - 2007 Integra LifeSciences Corporation. All Rights Reserved.
Newdeal® is a registered trademark of Newdeal, an Integra LifeSciences Corporation.
Integra Wave Logo is a trademark of Integra LifeSciences Corporation.

Terms of Use • Privacy Policy