  Mid & HindFoot Solutions
  Product Details
  Indications for Use/Contraindications
  Surgical Technique (4.0 mm)
 
 
    Determining Length of Screw
    Use of Drill
    Use of the Tap
    Screw Insertion
    Optional Dynamic Compression
    Optional Translation of Screw Body

Surgical Technique (6.5 mm)
Previous | Next Next


Positioning the Pin

  • A 2.5 mm diameter graduated K-wire, 200 mm long (115225ND) is introduced into the site where the I.CO.S.® 6.5 mm screw diameter is to be implanted
  • The K-wire is positioned by means of drilling guides and soft tissue protector (119541ND, 119551ND and 119552ND)
  • Failure to use the drilling guide and protecting sleeve when inserting the K-wire may result in soft tissue entrapment
  • Care should be taken when the K-wire is inserted to respect neighboring blood vessels, nerves and tissues
  • The depth of inseretion of the K-wire depends upon the surgeons will to achieve a mono or bi-cortical osteosynthesis