index  
coursenewsroomcompanycontactlogin
 
 
  Mid & HindFoot Solutions
  Product Details
  Indications for Use/Contraindications
 
 
    Determining Length of Screw
    Use of Drill
    Use of the Tap
    Screw Insertion
    Optional Dynamic Compression
    Optional Translation of Screw Body
  Surgical Technique (6.5 mm)

Surgical Technique (4.0 mm)
Previous | Next Next


Positioning the Pin

  • A 1.6 mm diameter, 15 cm long, with single trocar point (reference 115116ND) is introduced into the site where the I.CO.S® screw is to be implanted
  • The pin is positioned by means of a drilling guide (reference 119505ND). Failure to use the drilling guide when inserting the pin may result in tissues entrapment.
  • Care should be taken when pin is inserted to respect neighboring blood vessels, nerves and tissues
  • The depth of insertion of the pin depends upon the surgeon's wish to achieve a mono or bi-cortical osteosynthesis